Periodical calibration If the sample gas flow is determined b
y differential flow measurement the flow meter or the flow measurement instrumentation shall be calibrated by one of the following proced
ures, such that the probe flow GSE into the tunnel fulfils the accuracy requirements of Appendix I section 2.4: The flow meter for GDILW is connected in series to the flow meter for G TOTW , the difference between the two flow meters is calibrated for at least 5 set points with flow values equally spaced between the lowest GDILW value used
...[+++]during the test and the value of G TOTW used during the test The dilution tunnel may be bypassed.Periodieke kalibrering Wanneer de bemonsteringsgasstroom door middel van stroomverschilmeting wordt bepaald, moet de stroommeter of het stroommeetinstrumentarium volgens één van de volgende procedures worden gekali
breerd, om te zorgen dat de bemonsterde uitlaatgasmassastroom GSE in de tunnel voldoet aan de nauwkeurigheidseisen van punt 2.4 van aanhangsel 1: De stroommeter voor GDILW wordt in serie geplaatst met de stroommeter voor G TOTW ; het verschil tussen beide stroommeters wordt voor ten minste vijf instelpunten gekalibreerd, waarbij de stroomwaarden liggen op gelijke afstanden tussen de laagste waarde voor GDILW tijdens de test en
...[+++] de waarde voor G TOTW tijdens de test.